White Paper
Implementing NSQHS Infection Control Standards for Sterilisers and Washer-Disinfectors in Australian Healthcare
October 2026
White Paper
Implementing NSQHS Infection Control Standards for Sterilisers and Washer-Disinfectors in Australian Healthcare
October 2026
Disclaimer
This content is provided for information only. The authors make no representation or warranty regarding the accuracy, completeness or currency of the content. No information in this whitepaper should be construed as medical advice. Readers should seek appropriate professional guidance before acting on any information contained in this document. The authors expressly disclaim all liability for any direct or indirect loss or damage arising from the use of or reliance on this information.
Executive Summary
This paper sets out how to meet the reprocessing requirements of the National Safety and Quality Health Service (NSQHS) Standards for steam sterilisers, instrument washer-disinfectors (WDs) and bedpan WDs, using the requirements of AS 5369:2023.
Standards: NSQHS Action 3.17 requires reprocessing consistent with national and international standards and manufacturers' guidelines, traceability of critical and semi-critical devices to the patient and procedure, and processes for novel and emerging infections. AS 5369:2023 is the applicable national standard.
Records: Keep the records listed in AS 5369:2023, check the process record at the end of every cycle, and retain records for the period set by the relevant authority or, in its absence, facility policy.
Maintenance: Plan preventive maintenance from the equipment's documented procedures, perform it by a competent person, specify a maximum permitted delay, perform an annual electrical safety check, and record each activity with the details AS 5369:2023 requires.
Validation: Perform installation, operational and performance qualification (IQ, OQ and PQ), repeat OQ on the specified triggers, requalify annually, and apply the routine monitoring frequencies of AS 5369:2023.
Improvement: Audit regularly, control nonconforming product, investigate failures to root cause, and maintain a business continuity plan.
Introduction
Sterilisers and WDs deliver the cleaning, disinfection and sterilisation steps of reprocessing. NSQHS accreditation assesses whether these processes conform to current standards and whether the organisation can prove it through records. This paper converts the requirements into record templates, maintenance and validation regimes and monitoring frequencies for technical, sterilising services and infection prevention staff.
Overview of NSQHS Infection Control Standards
Action 3.17: The Preventing and Controlling Infections Standard requires that, when reusable equipment and devices are used, the health service organisation has:
AS 5369:2023: Published 15 December 2023, superseding AS/NZS 4187:2014 and AS/NZS 4815:2006. It applies wherever reusable medical devices are reprocessed and requires equipment to conform to ISO 15883 for WDs, EN 285 for large and EN 13060 for small steam sterilisers, and ISO/TS 22421.
Advisory AS24/01: The Australian Commission on Safety and Quality in Health Care (ACSQHC) required a gap analysis against AS 5369:2023 by 30 June 2025 and a compliance plan with timeframes aligned with national or international standards, manufacturers' guidelines, the current Australian Guidelines for the Prevention and Control of Infection in Healthcare and state and territory requirements. AS 5369:2023 sets no compliance deadline. Assessors check evidence that the plan is being implemented.
Novel and emerging infections: Define in advance the controls to apply, e.g. single-use devices, dedicated equipment or modified processes, and validate any modified process before use. For bedpan WDs, thermal disinfection does not inactivate bacterial spores, including Clostridioides difficile, so outbreak controls for spore-forming organisms must address cleaning efficacy, not disinfection temperature.
Governance: Assign responsibility for reprocessing to a competent person, report audit results, failures and corrective actions to the infection prevention and quality committee, and integrate equipment risks into the organisation's risk register.
Compliant Record Keeping Practices for Sterilisation and Disinfection
Required records: AS 5369:2023 requires records of at least: purchasing of devices and equipment, equipment monitoring, cleaning, sterilisation and high-level disinfection processes, microbiological surveillance where applicable, cleaning of equipment and the facility, staff training and competency, rosters, maintenance, IQ, OQ and PQ, process deviations with corrective and preventive actions, and recalls. Records must be controlled, retrievable and retained for not less than the period set by the relevant authority or, in its absence, facility policy.
Steriliser cycle records: Check the process record at the end of each cycle against the validated specification. Record the steriliser, cycle number, programme, date, operator, load contents, physical parameters, indicator results, BI results where used, release decision and the person releasing the load.
WD cycle records: Check the process record at the end of each cycle. Confirm water pressure, flow and spray action, cleaning agent dosage, and the temperature and time at disinfection temperature, or the A0 value, against the validated limits. Record the WD, programme, load, operator and outcome.
Traceability: AS 5369:2023 requires critical devices to be traceable at least to the individual cycle of a specified steriliser, and semi-critical devices undergoing high-level disinfection to the individual disinfection cycle. To meet Action 3.17, link those cycle records to the patient and procedure through load labels in the patient record or a scanned tracking system. Test the link with tracer audits.
Nonconformance records: Record every failed cycle, failed test, wet or damaged pack and recall, with the investigation, corrective action and effectiveness check.
Training records: Record training, competency assessment and reassessment for every person who operates, maintains or validates reprocessing equipment.
Template: Steriliser Cycle Log (Example)
| Date | Steriliser and Cycle | Operator | Load Contents | Physical Parameters | Indicators | Release Decision and Releasing Person |
|---|---|---|---|---|---|---|
| 3 Aug 2026 | CSSD steriliser 1, cycle 0412, 134 deg C [273 deg F] porous load | JD | Orthopaedic set 5, general sets 2 and 7 | Within validated limits, 134 deg C [273 deg F] held 4 min. Record attached | External indicators changed. Internal Type 5 indicators passed | Released. KL |
| 3 Aug 2026 | CSSD steriliser 1, cycle 0413, 134 deg C [273 deg F] porous load | JD | Implant set, spinal cages | Within validated limits. Record attached | External and internal indicators passed. BI in process challenge device, lot 123, no growth | Quarantined until BI result, then released. KL |
| 3 Aug 2026 | CSSD steriliser 2, cycle 0207, 134 deg C [273 deg F] porous load | MP | General sets 3, 4 and 9 | Not within limits. Temperature fell to 132 deg C [270 deg F] during holding | Internal indicator failed | Rejected. Load reprocessed. Steriliser withdrawn and fault reported. KL |
Example steriliser cycle log. Each row is one cycle. Adapt the fields to the tracking system in use.
Preventative Maintenance of Sterilisers and Washer-Disinfectors
Planned maintenance: AS 5369:2023 requires preventive maintenance of all equipment to be planned and carried out by a competent person in accordance with the documented procedures provided with the equipment, with the procedure and frequency of each activity specified. All reprocessing equipment requires an annual electrical safety check. Base maintenance and recalibration intervals on maintenance and calibration history, with no more than 12 months between recalibrations.
Return to use: Do not reprocess devices on equipment until the specified maintenance has been completed and documented. Specify the maximum permitted delay for any scheduled maintenance.
Daily user checks: Perform the checks in the IFU before first use: chamber and door seal condition, strainers and spray arms, cleaning agent levels, printer or data record function, and the daily air removal test on vacuum sterilisers.
Calibration: Calibrate timers, gauges and temperature monitoring devices at specified intervals with test instruments whose calibration is traceable to national standards.
Pressure equipment: Steam sterilisers and steam generators are pressure equipment. Confirm the AS 4343 hazard level, whether item registration applies under Schedule 5 of the WHS Regulations, and the in-service inspection regime to AS/NZS 3788. File inspection certificates with the maintenance records.
Repairs: Withdraw faulty equipment, label it out of service and record the fault. After repair, perform OQ before return to use. AS 5369:2023 allows OQ after repair to consist of performance tests, e.g. leak rate, air removal and thermometric tests for a steriliser.
Bedpan WDs: Include strainers, spray nozzles, steam generator, door seal, dosing system and water supply in each planned visit. Verify the thermal disinfection A0 value with independent sensors at requalification.
Template: Preventative Maintenance Record
The fields follow the minimum maintenance record content of AS 5369:2023.
| Date and Reason | Equipment (Model, Serial, Location) | Work Performed and Parts Replaced | Performed By | Tests After Work | Released By | Next Due |
|---|---|---|---|---|---|---|
| 10 Jul 2026. Planned annual service | Steam steriliser, serial 4471, CSSD | Door seal replaced. Temperature sensor calibrated, error of 0.2 deg C [0.4 deg F] corrected. Vacuum pump serviced. Safety valve tested | Service contractor, J. Smith | Leak rate, air removal and thermal penetration tests passed | CSSD manager | Jul 2027 |
| 15 Jul 2026. Planned six-monthly service | Bedpan WD, serial 2210, Ward 5 sluice | Steam generator descaled. Strainers and nozzles cleaned. Door seal inspected | Biomedical services, T. Lee | Thermometric test: A0 at all sensors not less than the programmed A0 of 600 | Nurse unit manager | Jan 2027 |
| 5 Aug 2026. Repair, cycle aborted on steam generator fault | Steam steriliser, serial 3982, theatre suite | Steam generator heating element replaced | Biomedical services, S. Khan | OQ performance tests: leak rate, air removal and thermometric tests passed | CSSD manager | Planned service Dec 2026 |
Example maintenance record covering planned services and an unplanned repair.
Validation Protocols and Ongoing Performance Monitoring
Qualification
Routine Monitoring
| Test | Equipment | Frequency | Action on Failure |
|---|---|---|---|
| Process record check | All sterilisers and WDs | Every cycle | Reject the load, reprocess, investigate |
| Air removal and steam penetration test (Bowie and Dick type) | Steam sterilisers with vacuum air removal | Daily, before first load | Withdraw steriliser, repair, repeat test |
| Leak rate test | Large steam sterilisers | Weekly with an automatic air detector, daily without | Withdraw steriliser, repair, repeat test |
| Air detector function test | Large steam sterilisers | Weekly | Withdraw steriliser, repair, repeat test |
| External chemical indicator | All sterilisers | Each pack | Do not release the pack |
| BI | Steam, VH2O2, EO sterilisers | Steam: set by the facility. VH2O2: per IFU. EO: every load | Recall the load, withdraw steriliser, investigate |
| Cleaning efficacy | Instrument WDs | Visual inspection of every item. Efficacy test daily and quarterly | Reprocess, check spray system and dosing, retest |
| Supply water hardness and chloride | Instrument WDs | Monthly, adjusted by results | Treat water, increase test frequency |
| Final rinse water conductivity, TVC and endotoxin | Instrument WDs | Conductivity and TVC monthly, endotoxin annually | Treat water, increase test frequency |
| Steriliser feedwater | Sterilisers with a dedicated steam generator | At IQ or OQ, then annually | Correct treatment, retest |
| Thermal penetration and drying effectiveness | Large and small steam sterilisers | Annually | Investigate, adjust load or cycle, revalidate |
| Thermometric disinfection test | Instrument and bedpan WDs | Annually, at requalification | Withdraw WD, repair, revalidate |
Routine monitoring frequencies from AS 5369:2023. Monthly TVC testing may be reduced to a minimum of annually after 12 months of results within specification.
BI frequency for steam: AS 5369:2023 leaves BI frequency for moist heat to the facility. Its informative example process uses a BI in a process challenge device at least weekly, preferably daily, and in every implant load, with implant loads quarantined until the BI result is known.
Chemical indicators: Select indicators to ISO 11140-1 by type: Type 1 process indicators on each pack, Type 2 for air removal tests, and Types 4 to 6 for internal monitoring where facility policy requires it.
Bedpan WDs: ISO 15883-3 requires the thermal disinfection stage to deliver an A0 value of at least 60, equivalent to 80 deg C [176 deg F] for 1 min. Program a higher A0 value, e.g. A0 600, equivalent to 90 deg C [194 deg F] for 1 min, where the infection prevention risk assessment requires it. Confirm the A0 value or temperature and time on the process record every cycle.
Protocol Example: Annual Requalification of a Steam Steriliser
Compliance Best Practices for Risk Mitigation and Continuous Quality Improvement
Conclusion
NSQHS Action 3.17 is met for sterilisers and WDs by operating them to AS 5369:2023 and proving it through records. Keep the required records and check every cycle record. Maintain equipment to documented procedures with a specified maximum delay and annual electrical safety checks. Qualify, monitor and requalify at the AS 5369:2023 frequencies. Link cycle records to patients for critical and semi-critical devices. Use audits, nonconformance control and trend analysis to correct failures, and keep the gap analysis and compliance plan required by Advisory AS24/01 current.
Sources