White Paper
Continuous Improvement Framework for Sterilisation Equipment Compliance Auditing in Australian Healthcare
October 2026
White Paper
Continuous Improvement Framework for Sterilisation Equipment Compliance Auditing in Australian Healthcare
October 2026
Disclaimer
This content is provided for information only. The authors make no representation or warranty regarding the accuracy, completeness or currency of the content. No information in this whitepaper should be construed as medical advice. Readers should seek appropriate professional guidance before acting on any information contained in this document. The authors expressly disclaim all liability for any direct or indirect loss or damage arising from the use of or reliance on this information.
Introduction
This paper sets out an audit and continuous improvement framework for sterilisers, washer-disinfectors and reprocessing processes in hospitals, aged care facilities and office-based practices. It defines the audit criteria, evidence, metrics, corrective action process and audit frequency, based on AS 5369:2023.
Requirements Basis
AS 5369:2023: Superseded AS/NZS 4187:2014 and AS/NZS 4815:2006, and applies to all facilities that reprocess reusable medical devices. It requires regular periodic audits to confirm that its requirements are met, documented findings, corrective action where required, and review of each corrective action for effectiveness. It sets no implementation timeframe and requires a risk-based approach.
NSQHS Standards: Hospitals and day procedure services are assessed against Action 3.17 of the Preventing and Controlling Infections Standard, which covers reprocessing of reusable equipment, instruments and devices. Australian Commission on Safety and Quality in Health Care Advisory AS24/01 requires a gap analysis against AS 5369:2023 and a compliance plan.
Other Settings: Residential aged care is assessed against the Aged Care Quality Standards, which include infection prevention and control. Registered dental practitioners must comply with the Dental Board of Australia infection control guidelines. Both settings apply AS 5369:2023 where devices are reprocessed.
Crosswalk: Map every audit item to the AS 5369:2023 clause it verifies. Update the audit tool when the standard or facility procedures change.
Audit Checklist
Master Checklist: Build one master checklist covering the full reprocessing cycle and the equipment quality system. Remove items that do not apply to a site and add site-specific items, e.g. bedpan washer-disinfector records for hospitals and aged care, and handpiece reprocessing for dental practices. Audit records and direct observation, not policy documents alone.
| Audit Area | Evidence to Check | AS 5369:2023 Clause |
|---|---|---|
| Equipment conformance | Conformance to ISO 15883, EN 285 or EN 13060 and ISO/TS 22421. ARTG entry for each machine. | 2.5.2, 4.3.3 |
| Validation | Approved IQ, OQ and PQ reports. Requalification within 12 months. Assessment and requalification after each change or major repair. | 7, 10.4, 10.5 |
| Steriliser routine monitoring | Daily air removal and steam penetration test results. Leak rate test results at the required frequency. Process record checked for every cycle. External chemical indicator on every pack. | 8.7 |
| Washer-disinfector routine monitoring | Process record checked for every cycle, including disinfection temperature and time. Dosing volume verification. Cleaning efficacy results. | 8.2 |
| Water and steam quality | Pre-installation water tests. Routine water results at the required frequencies. Feedwater or condensate tests within 12 months. Steam quality tests after installation, relocation or supply change. | 7.2.3, 8.2 |
| Calibration | Calibration status label on each instrument. Calibration reports with the certification number of the reference instrument. Recalibration within the specified interval. | 2.5.4, 10.2 |
| Maintenance | Preventive maintenance completed to schedule and within the maximum permitted delay. Complete maintenance records. Annual electrical safety check. | 10.3 |
| Release and nonconformance | Release against the specified criteria. Quarantine and investigation of nonconforming items, e.g. wet packs and damaged packaging. | 9.2, 2.6.2 |
| Traceability and recall | Traceability of each critical device to its cycle and steriliser. Recall procedure. Recall reports. | 2.5.3, 2.6.3 |
| Reprocessing environment | Segregation of clean and dirty activities. Unidirectional workflow. Process map. Finishes and dedicated sinks. | 5.6 |
| Staff | Orientation and training program. Competency assessments within the defined interval. Training and roster records. | 2.4.3, 2.3.3 |
| Contracts | Agreements for outsourced reprocessing, maintenance and testing that define responsibilities and conformance to AS 5369:2023. Evidence of the provider's accreditation or audit. | 2.4.5 |
Hospitals: Audit the full CSSD workflow from point-of-use collection to sterile storage, all steam and low temperature sterilisers, washer-disinfectors, water treatment, and ward bedpan washer-disinfectors.
Aged Care: Audit bedpan washer-disinfector process records and maintenance, cleaning of commodes and shared equipment, storage of disinfected items, and contracts for any outsourced sterilisation.
Dental and Office-Based Practices: Audit small steriliser routine tests, cycle records, packaging and labelling, steriliser feedwater, maintenance and calibration records, and the reprocessing room layout.
Frequency: AS 5369:2023 does not set an audit frequency. Set it by risk and volume, e.g. monthly in a CSSD and quarterly in aged care and office-based practices. Review every process record at release in all settings. Increase frequency after a failed audit, an incident or a process change.
Metrics and Trend Analysis
Data Capture: Record each audit item as compliant, noncompliant or not applicable in a structured register so that results can be compared between periods, departments and sites.
Metrics: Define each metric with its calculation and target.
Audit compliance rate, as compliant items divided by applicable items.
Failed steriliser cycles per 100 cycles, by machine.
Failed air removal and leak rate tests per month, by machine.
Wet packs per 100 loads.
Nonconforming items per 1,000 items processed, by cause.
Preventive maintenance and calibration tasks past due.
Water quality exceedances per quarter, by parameter.
Repeat findings, and median days to close corrective actions.
Analysis: Plot each metric on a run chart. Investigate any metric outside its target, any adverse trend over three consecutive periods, and any item that fails in consecutive audits. Stratify by machine, shift, department and site to locate the source.
Reporting: Report results to the infection prevention or quality committee at least quarterly, and to reprocessing staff after each audit. Compare results between departments and sites within the organisation.
Improvement Cycle: Use a plan, do, study, act cycle. Plan an intervention from the data, implement it, re-audit the affected items, and either standardise the change or revise it.
Corrective Action
Containment: Act on critical findings during the audit. Quarantine items released without meeting the release criteria, recall any that have been distributed, and remove from service any machine operating outside its specification.
Corrective Action Register: Log each finding with its risk rating, owner, due date and status. For each nonconformity, AS 5369:2023 requires identification of its nature, an action plan, documentation of the action, evaluation of effectiveness, and additional action where required.
Root Cause Analysis: Analyse repeat and systemic findings with structured methods, e.g. five whys or cause and effect diagrams. Consider equipment condition, utilities, procedures, staffing, training and workload. Record preventive actions with the corrective actions.
Process and Document Change: Where the fix changes a procedure, update the procedure under document control, train affected staff, and add the change to the next audit. Where the change affects a validated process, assess it and repeat IQ, OQ or PQ as required.
Verification: Close a corrective action only after a follow-up audit or record review confirms it has been effective.
Escalation: Escalate findings that require capital, staffing or service changes to the governance committee with the audit data, failure history and risk rating.
Reporting Culture: Report findings against processes and systems rather than individuals, and encourage staff to report deviations and propose fixes.
Training
Requirements: AS 5369:2023 requires a formal orientation and training program, trained and competent staff, and competency reassessment at intervals defined by the facility. Its guidance recommends annual training in infection prevention and control and in occupational exposure procedures.
Modules: Structure training in modules that follow the workflow, each with learning objectives and an assessment.
Standards, NSQHS requirements and the facility audit criteria.
Cleaning, washer-disinfector operation, process record checks and water quality.
Inspection, assembly, packaging and labelling.
Steriliser operation, routine tests, chemical indicators of Types 1 to 6 to ISO 11140-1, biological indicators and load release.
Sterile storage, transport and handling.
Documentation, traceability, quarantine and recall.
Work health and safety and exposure management.
Self-audit and response to audit findings.
Link to Audit: Deliver targeted training for each audit finding attributed to practice, then verify by observation. Use recall and failed-cycle drills to assess response to non-routine events. Record all training and assessments.
Qualifications: Specify HLT37015 Certificate III in Sterilisation Services for technicians and HLT47015 Certificate IV for team leaders where the employer requires a qualification.
Monitoring Tools
Instrument Tracking Systems: Scan device sets at each stage of reprocessing, link each set to its cycle record and release, and hold sets automatically when a step or test result is missing. Record nonconformances as quality events in the system and use its reports for audit evidence.
Equipment Data: Export cycle records electronically from sterilisers and washer-disinfectors to the tracking or records system. Confirm controller clocks are correct. Review cycle data weekly for parameter drift.
Electronic Audit Tools: Use audit software or forms that timestamp entries, attach photographs, standardise findings, score results and create corrective actions directly from noncompliant items. Schedule audits, calibration, maintenance and requalification with automatic reminders.
Dashboards: Display the defined metrics, overdue tasks and loads awaiting release on a dashboard reviewed daily by supervisors and quarterly by the governance committee.
Small Sites: Use structured electronic logs with mandatory fields and alerts for missed daily tests in place of paper logbooks.
Consumables: Track lot numbers and expiry dates of indicators and packaging, and set minimum stock levels so that monitoring is never interrupted.
Scaling by Risk
Consistent Outcomes: Apply the same acceptance criteria in every setting. Implementation may differ, e.g. an electronic tracking system in a CSSD and a structured log in a small practice, but each must provide the same evidence.
Tiered Model: Allocate audit effort by risk and volume.
Tier 1: CSSDs and other high-volume reprocessing. Frequent audits, dedicated auditors, electronic tracking and monthly metric review.
Tier 2: Day procedure units and larger practices. Regular audits, shared auditors and periodic external review.
Tier 3: Low-volume sites and outsourced services. Self-audit against the master checklist with periodic central audit.
Organisation Standards: Define requirements that apply at every site, e.g. requalification within 12 months, calibration within interval and traceability for critical devices.
Outsourced Reprocessing: Audit the contract, the provider's accreditation and audit evidence, indicator checks on receipt, and transport conditions.
Rescaling: Move a site to a higher tier when its risk increases, e.g. when it introduces complex surgical devices or a new reprocessing process. Rotate auditors between sites and share effective tools across the organisation.
Key Actions
References